Frequency (Autoranging)
Range
Resolution
Accuracy
9.999Hz
0.001Hz
99.99Hz
0.01Hz
+1.5% of rdg + 5 dgts
999.9Hz
0.1Hz
9.999kHz
1Hz
99.99kHz
10Hz
999.9kHz
100Hz
+1.2% of rdg + 3 dgts
9.999MHz
1kHz
+1.5% of rdg + 4 dgts
Sensitivity: >0.5V RMS while ≤1MHz;
Sensitivity: >3V RMS while >1MHz;
Overload protection: 250VDC or AC rms.
Duty Cycle
Range
Resolution
Accuracy
0.1%~99.9%
0.1%
+1.2% of rdg + 2 dgts
Pulse width: >100us, <100ms;
Frequency width: 5Hz – 150kHz
Sensitivity: >0.5V RMS
Overload protection: 250V DC or AC rms.
Temperature
Range
Resolution
Accuracy
-20
o
C~+760
o
C
1
o
C
-4
o
F~+1400
o
F
1
o
F
+3% of rdg
+5
o
C/9
o
F
Sensor: Type K Thermocouple
Overload protection: 250VDC or AC rms.
Diode Test
Test current
Resolution
Accuracy
0.3mA typical
1 mV
+10% of rdg + 5 dgts
Open circuit voltage: 1.5V DC typical
Overload protection: 250V DC or AC rms.
Audible Continuity
Audible threshold: Less than 150Ω
Test current: <0.3mA
Overload protection: 250V DC or AC rms.