7 software update, 8 how to proceed in case of repair – VEGA VEGAFLEX 81 Modbus and Levelmaster protocol User Manual
Page 68
68
9 Diagnostics and service
VEGAFLEX 81 • Modbus and Levelmaster protocol
41828-EN-130314
4
1
1
2
2
3
A
B
C
45 mm (1.77")
100 mm
(3.94")
25 mm (0.98")
50 mm
(1.97")
Fig. 36: Shortening the cable probe
A Gravity weight - cable ø 4 mm
B Gravity weight - cable ø 2 mm
C Centering weight - cable ø 2 mm
1 Pins
2 Thread M8 for eye-bolt
3 Fixing screw - centering weight
9.7 Software update
The following components are required to update the sensor soft-
ware:
•
Sensor
•
Voltage supply
•
Interface adapter VEGACONNECT 4
•
PC with PACTware
•
Current sensor software as file
You can find the actual sensor software as well as detailed infor-
mation of the procedure under "www.vega.com/downloads" and
"Software".
You can find information about the installation in the download file.
Caution:
Instruments with approvals can be bound to certain software ver-
sions. Therefore make sure that the approval remains effective with a
software update.
You can find detailed information on www.vega.com/downloads and
"Approvals".
9.8 How to proceed in case of repair
You can find a repair form as well as detailed information on how to
proceed under www.vega.com/downloads and "Forms and certifi-
cates".
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.
If a repair is necessary, please proceed as follows:
•
Print and fill out one form per instrument