2
Mounting
Take note of the operating instructions of the corresponding sensor
MINITRAC and the source holder VEGASOURCE.
Take note of the following mounting instructions:
l
Mount the bracket first, then the sensor and the source container
l
The arrow cutout in the clamp (source container side) and the
transport lug of the source container must point in the same
direction (A) after mounting
l
Make sure that the two clamps (5 and 8) of the bracket are in
parallel. Do this by measuring the lateral distances between the
clamps
l
To avoid injuries, shorten the threaded rods (1) of the brackets to a
suitable length after mounting
Mount the bracket according to the following assembly drawing:
7
6
7
6
5
3
8 1
A
4
2
Fig. 3: Mounting bracket with horizontally mounted sensor
1
Threaded rod M10 x 620 mm (M10 x 42.41 in) - 4 pcs.
2
Hexagon nut M16 - 4 pcs.
3
Hexagon nut M10 - 16 pcs.
4
Washer for M16 - 4 pieces
5
Clamp - Source holder side (VEGASOURCE 31) - 1 piece
6
Washer for M10 - 8 pieces
7
Hexagon screw M8 - 2 pieces
8
Clamp - Sensor side (MINITRAC) - 1 piece
A
Arrow cutout of the clamp and the lug point in the same direction
Operating instructions
Mounting brackets for
horizontal mounting
Horizontal sensor
mounting
Mounting brackets KV 31 • Source holder SE
5
2 Mounting
41407
-
EN