VEGA WEIGHTRAC 31 4 … 20 mA_HART - four-wire User Manual
Page 53
53
7 Setup with PACTware
WEIGHTRAC 31 • 4 … 20 mA/HART - four-wire
42374-EN-131119
You can select further options in PACTware.
A detailled description of the functions is available in the PACTware
online help.
When a reset is carried out, all settings (with only a few exceptions)
are reset. The exceptions are: PIN, language and HART mode.
The following reset functions are available:
Basic settings: Restores the parameter settings at the time of ship-
ment Ex factory including the order-specific settings. The measured
value memory is deleted.
Factory settings: Resets parameter settings including special
parameters to the default values of the respective instrument. The
measured value memory and the order-specific settings are deleted.
Peak values: Resets the parameter settings in the menu item "Setup"
to the default values of the respective instrument. Order-specific
settings remain but are not taken over into the current parameters.
The measured value memory as well as the event memory remain
unaffected.
The following table shows the default values of the instrument. The
values apply for the application "Mass flow". The application must be
selected first.
Depending on the instrument version, not all menu items may be
available or they may be differently assigned:
Menu
Menu item
Default value
Setup
Measurement
loop name
Sensor
Isotope
Cs-137
Application
Mass flow
Background ra-
diation
0 ct/s
Unit of the pro-
cess value
m
Temperature unit °C
Damping
60 s
Real value cor-
rection
0
Current output
mode
4 … 20 mA, < 3.6 mA
Current output
Min./Max.
Min. current 3.8 mA, max. current
20.5 mA
X-ray alarm
Modulated measuring current
Reference value
- Relay
None
Block adjustment Released
Further settings in PACT-
ware
Additional adjustments/
Reset