VEGA VEGABAR 52 4 … 20 mA User Manual
Page 46
46
10 Supplement
VEGABAR 52 • 4 … 20 mA
36716-EN-130321
0,15 %
-0,15 %
-40°C
-20°C
20°C
40°C
60°C
80°C
Fig. 19: Thermal change, current output
Long-term stability (according to DIN 16086 and IEC 60770-1)
Applies to digital HART interface as well as to analogue current output 4 … 20 mA under refer-
ence conditions. Specifications refer to the set span. Turn down (TD) is the relation nominal meas-
uring range/set span.
Long-term drift of the zero signal:
Ʋ For one year
< 0.05 % x TD
Ʋ For five years
< 0.1 % x TD
Ʋ For ten years
< 0.2 % x TD
Ambient conditions
Ambient, storage and transport temperature
Ʋ Standard version
-40 … +80 °C (-40 … +176 °F)
Ʋ Versions IP 66/IP 68 (1 bar) and IP 68
(25 bar), connection cable PUR
-20 … +80 °C (-4 … +176 °F)
Ʋ Version IP 66/IP 68 (1 bar), connec-
tion cable PE
-20 … +60 °C (-4 … +140 °F)
Process conditions
The specifications of the pressure stage and product temperature are used as an overview. The
specifications on the type label are applicable.
Pressure stage, process fitting
Ʋ Thread 316L, depending on connec-
tion
PN 10, PN 60, PN 160
Ʋ Thread Aluminium
PN 25
Ʋ Thread PVDF
PN 10
Ʋ Hygienic fittings 316L, depending on
connection
PN 6, PN 10, PN 25, PN 40 (PN 40 only with DRD and
DIN 11851)
Ʋ Flange 316L
PN 16, PN 40,150 lbs, 300 lbs, 600 lbs
Ʋ Flange with extension 316L
without PN specification, PN 16, PN 40 or 150 lbs,
300 lbs, 600 lbs
Ʋ Flange flattened on both sides 316L
PN 10
Ʋ Flange PVDF
PN 16