beautypg.com

VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART - four-wire User Manual

Page 68

background image

68

11 Supplement

VEGAPULS 61 • 4 … 20 mA/HART - four-wire

36500-EN-121011

Wire cross-section (spring-loaded terminals)

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Ʋ Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

Indicating and adjustment module
Voltage supply and data transmission

through the sensor

Indication

LC display in dot matrix

Measured value indication

Ʋ Number of digits

5

Ʋ Size of digits

W x H = 7 x 13 mm

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the sensor without cover IP 40

Materials

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Integrated clock
Date format

Day.Month.Year

Time format

12 h/24 h

Time zone Ex factory

CET

Measurement electronics temerature
Resolution

1 °C (1.8 °F)

Accuracy

±1 °C (1.8 °F)

Voltage supply
Operating voltage

Ʋ Version for low voltage

9.6 … 48 V DC, 20 … 42 V AC, 50/60 Hz

Ʋ Version for mains voltage

90 … 253 V AC, 50/60 Hz

Interpolation protection

Integrated

Max. power consumption

4 VA; 2.1 W

Electrical protective measures
Protection rating

IP 66/IP 67

Overvoltage category

III

Protection class

I

Approvals
Instruments with approvals can have different technical data depending on the version.
For that reason the associated approval documents of these instruments must be carefully noted.

They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as