beautypg.com

VEGA POINTRAC 31 8_16 mA_HART - four-wire User Manual

Page 47

background image

47

11 Supplement

POINTRAC 31 • 8/16 mA/HART - four-wire

39411-EN-131119

Relay output
Output

Relay output (SPDT), floating spdt

Switching voltage

Ʋ Min.

10 mV

Ʋ Max.

253 V AC, 253 V DC

Switching current

Ʋ Min.

10 µA

Ʋ Max.

3 A AC, 1 A DC

Breaking capacity

Ʋ Min.

50 mW

Ʋ Max.

750 VA AC, 40 W DC
If inductive loads or stronger currents are switched

through, the gold plating on the relay contact surface will

be permanently damaged. The contact is then no longer

suitable for switching low-level signal circuits.

Contact material (relay contacts)

AgNi or AgSnO and Au plated

Transistor output
Output

Floating transistor output, permanently shortcircuit-proof

Load current

< 400 mA

Voltage loss

< 1 V

Switching voltage

< 55 V DC

Blocking current

< 10 µA

Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1

Ʋ Temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Repeatability

≤ 0.5 %

Deviation with bulk solids

The values depend to a great extent on the application.

Binding specifications are thus not possible.

Deviation under EMC influence

≤ 1 %

Variables influencing measurement accuracy
Specifications apply also to the current output
Temperature drift - Current output

±0.03 %/10 K relating to the 16 mA span max. ±0.3 %

Deviation on the current output through

analogue/digital conversion

<±15 µA

Deviation on the current output due to

strong, high frequency electromagnetic

interference acc. to EN 61326

<±150 µA