10 Supplement
VEGABAR 54 • 4 … 20 mA/HART - climate compensated
36727-EN-130321
Ʋ Transport cover
PFA
Weight approx.
0.8 … 8 kg (1.764 … 17.64 lbs), depending on process
fitting
Max. torques
Ʋ Screws for PMC 1", PMC 1¼"
2 Nm
Ʋ Screws for PMC 1½"
5 Nm
Ʋ G½ A, G¾ A
30 Nm
Ʋ G1 A, M30 x 1.5
50 Nm
Ʋ G1" for PASVE
100 Nm
Output variable
Output signal
4 … 20 mA/HART
HART output values according to HART standard 5.0
Ʋ Primary Value
Pressure
Ʋ Secondary Value
Temperature
9)
Signal resolution
1.6 µA
Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Max. output current
22 mA
Load
see load diagram under Power supply
Met NAMUR recommendation
NE 43
Dynamic behaviour output
Run-up time approx.
10 s
90 %
100 %
10 %
t
t
T
A
t
t
S
2
1
Fig. 24: Sudden change of the process variable. t
T
: dead time; t
A
: rise time; t
S
: jump response time
1 Process variable
2 Output signal
Dead time
≤ 150 ms
Rise time
≤ 100 ms (10 … 90 %)
Step response time
≤ 250 ms (ti: 0 s, 10 … 90 %)
9)
Depending on the respective instrument and the process fitting, the temperature can deviate from the real
process temperature.