VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) HART - integrated accumulator User Manual
Page 63

63
11 Supplement
VEGAPULS 61 • HART - integrated accumulator
40792-EN-121011
Vibration resistance
Ʋ With adapter flange
2 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ with mounting strap
1 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Shock resistance
100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
Indicating and adjustment module
Voltage supply and data transmission
through the sensor
Indication
LC display in dot matrix
Measured value indication
Ʋ Number of digits
5
Ʋ Size of digits
W x H = 7 x 13 mm
Adjustment elements
4 keys
Protection rating
Ʋ unassembled
IP 20
Ʋ mounted into the sensor without cover IP 40
Materials
Ʋ Housing
ABS
Ʋ Inspection window
Polyester foil
Indicating and adjustment elements
Indicating elements
Ʋ Green LED in the supply room
Indication of the charging process
Ʋ Yellow LED in the supply room
Indication of the charging condition
Adjustment elements
Ʋ Rotary switch in the supply room
Selection of the mode
Ʋ Button outside on the housing
Switching on and off
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone Ex factory
CET
Measurement electronics temerature
Resolution
1 °C (1.8 °F)
Accuracy
±1 °C (1.8 °F)
External battery charger
Mains voltage
100 … 240 V AC
Output voltage
24 V DC
Output current (shortcircuit proof) max. 500 mA
Load current limitation
70 mA