11 Supplement
VEGAPULS 63 • 4 … 20 mA/HART - four-wire
36512-EN-130228
Ʋ QV (Fourth Value)
Scaled measured value
Resolution, digital
< 1 mm (0.039 in)
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
Ʋ Temperature
+18 … +30 °C (+64 … +86 °F)
Ʋ Relative humidity
45 … 75 %
Ʋ Air pressure
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Installation reference conditions
Ʋ Min. distance to installations
> 200 mm (7.874 in)
Ʋ Reflector
Plane plate reflector
Ʋ False reflections
Biggest false signal, 20 dB smaller than the useful signal
Deviation with liquids
See following diagrams
0,5 m (1.6 ft)
10 mm (0.394 in)
- 10 mm (- 0.394 in)
0
2 mm (0.079 in)
- 2 mm (- 0.079 in)
1 2
3
Fig. 43: Deviation under reference conditions
1 Reference plane
2 Antenna edge
3 Recommended measuring range
Repeatability
≤ ±1 mm
Variables influencing measurement accuracy
Specifications apply to the HART signal and the current output
Temperature drift - Digital output
±3 mm/10 K relating to the max. measuring range or
max. 10 mm
Additional deviation through electromag-
netic interference acc. to EN 61326
< ±50 mm
Specifications apply also to the current output
Temperature drift - Current output
±0.03 %/10 K relating to the 16 mA span max. ±0.3 %
Deviation on the current output by ana-
logue/digital conversion
< ±15 µA
Deviation on the current output due to
strong, high frequency electromagnetic
fields acc. to EN 61326
< ±150 µA