beautypg.com

VEGA VEGABAR 55 Foundation Fieldbus - climate compensated User Manual

Page 50

background image

50

10 Supplement

VEGABAR 55 • Foundation Fieldbus - climate compensated

39443-EN-130321

0,15 %

-0,15 %

-40°C

-20°C

20°C

40°C

60°C

80°C

Fig. 27: Thermal change, current output

Long-term stability (according to DIN 16086 and IEC 60770-1)
Applies to digital HART interface as well as to analogue current output 4 … 20 mA under refer-

ence conditions. Specifications refer to the set span. Turn down (TD) is the relation nominal meas-

uring range/set span.
Long-term drift of the zero signal:

Ʋ For one year

< 0.05 % x TD

Ʋ For five years

< 0.1 % x TD

Ʋ For ten years

< 0.2 % x TD

Ambient conditions
Ambient, storage and transport temperature

Ʋ Standard version

-40 … +80 °C (-40 … +176 °F)

Ʋ Versions IP 66/IP 68 (1 bar) and IP 68

(25 bar), connection cable PUR

-20 … +80 °C (-4 … +176 °F)

Ʋ Version IP 66/IP 68 (1 bar), connec-

tion cable PE

-20 … +60 °C (-4 … +140 °F)

Process conditions
The specifications of the pressure stage and product temperature are used as an overview. The

specifications on the type label are applicable.
Pressure stage, process fitting

Ʋ Thread 316L, depending on connec-

tion

PN 25, PN 60

Ʋ Thread Aluminium

PN 25

Ʋ Hygienic fittings 316L, depending on

connection

PN 6, PN 10, PN 25, PN 40 (PN 40 only with DRD and

DIN 11851)

Ʋ Flange 316L

PN 16/PN 40 or 150 lb/300 lb

Product temperature

Ʋ Standard

-12 … +140 °C (+10 … +284 °F)

Ʋ with extenation, extended thread or

Clamp 2½"

-12 … +140 °C (+10 … +284 °F)

Ʋ with cooling element

-12 … +180 °C (+10 … +356 °F)