beautypg.com

4 software update – VEGA VEGAFLEX 67 (-20…+250°C) 4 … 20 mA_HART four-wire User Manual

Page 38

background image

I

n Ex applications only one instrument and one electronics module with

respective Ex approval may be used.

I

f there is no electronics module available on site, one can be ordered

from the VEGA agency serving you.

T

he new electronics module must be loaded with the settings of the

sensor. These are the options:

l

A

t the factory by VEGA

l

O

r on site by the user

I

n both cases, the sensor serial number is necessary. The serial

numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.

Information:
W

hen loading on site, first of all the order data must be downloaded

from the Internet (see operating instructions manual "Electronics
module
").

T

he electronics modules are adapted to the respective sensor and

distinguish also in the signal output or power supply.

E

lectronics module FX-E.67V suitable for VEGAFLEX 67 - 4 … 20 mA/

HART

four-wire:

l

FX-E.67VX (X =

without approvals)

8

.4 Software update

T

he software version of VEGAFLEX 67 can be determined as follows:

l

on the type label of the electronics

l

via the indicating and adjustment module

l

via PACTware

Y

ou can view all software histories on our website www.vega.com.

M

ake use of this advantage and get registered for update information

via e-mail.

T

he following components are required to update the sensor software:

l

S

ensor

l

V

oltage supply

l

VEGACONNECT

l

PC

with PACTware

l

C

urrent sensor software as file

A

t "www.vega.com/downloads" go to "Software". Select under "plics

sensors and instruments", "Firmware updates" the respective instru-
ment series and software version. Load the zip file via the right mouse
key with "Save target as" e.g. on the desktop of your PC. Move with the
right mouse key to the folder and select "Extract all". Save the
extracted files, for example on the desktop.

Sensor serial number

Assignment

4

… 20 mA/HART four-

wire

Load sensor software to
PC

38

VEGAFLEX

67 • 4 … 20

mA/HART - four-wire

8 M

aintenance and fault rectification

32308

-EN