VEGA VEGAFLEX 67 (-40…+150°C PFA insulated) Foundation Fieldbus User Manual
Page 40
-
Rod: ø 10 mm (0.394 in)
approx. 620 g/m (6.7 oz/ft)
-
Cable: ø 4 mm (0.157 in)
approx. 80 g/m (0.86 oz/ft)
-
Gravity weight
325 g (11.5 oz)
Insulation thickness
-
Rod version
2 mm (0.08 in)
-
Cable version
0.75 mm (0.03 in)
On the gravity weight and beginning of the cable:
2 mm (0.08 in)
Probe length L (from seal surface), cannot be shortened
-
Rod: ø 10 mm (0.394 in)
up to 4 m (13.12 ft)
-
Trimming accuracy - rod
< 1
mm (0.039 in)
-
Cable: ø 4 mm (0.157 in)
up to 32 m (105 ft)
-
Trimming accuracy - cable
±0
.05 %
Lateral load - rod: ø 10 mm (0.394 in)
4 Nm (3 lbf ft)
Max. tensile load with cable: ø 4 mm
(0.157 in)
2 KN (450 lbf)
Input variable
Measured variable
Level of liquids
Min. dielectric figure (lower medium)
by 10 higher than the ε
r
of the upper medium
Min. layer thickness (upper medium)
> 100
mm (3.937 in)
The measured value deviates from the actual level
until the min. layer thickness is reached.
Max. layer thickness (upper medium)
-
Dielectric figure (upper medium) 10
6 m (19.69 ft)
-
Dielectric figure (upper medium) 5
10 m (32.81 in)
-
Dielectric figure (upper medium) 2
20 m (65.61 in)
Min. dielectric figure (upper medium) with
rod, cable version
ε
r
> 1
.6
40
VEGAFLEX
67 • Foundation Fieldbus
10 Supplement
32306
-EN
-120329