beautypg.com

VEGA VEGACAL series 60 User Manual

Page 8

background image

4

Loosen the two holding screws with a screwdriver (Torx size T 10
or Phillips size 4)

2

1

Fig. 3: Loosening the holding screws

1

Electronics module

2

Screws (2 pcs.)

5

Pull the previous electronics out by holding the opening levers.

6

Insert the new electronics module carefully.

7

Insert the two hold screws and tighten them

8

Connect the connection cables according to the operating
instructions manual of the respective sensor

9

Screw the housing cover on

The electronics exchange is now finished.

As a rule, an exchange of electronics must be documented internally
when Ex applications are involved.