6 Setup
VEGASWING 63 • - NAMUR
29231-EN-130409
6 Setup
6.1 General information
The figures in brackets refer to the following illustrations.
With plastic housings, the switching condition of the electronics can
be checked when the housing cover is closed (control lamp). With the
basic setting, products with a density > 0.7 g/cm³ (0.025 lbs/in³) can
be detected. For products with lower density, the switch must be set
to > 0.5 g/cm³ (0.018 lbs/in³).
On the electronics module you will find the following indicating and
adjustment elements:
•
Signal lamp (1)
•
DIL switch for characteristics reversal - min./max. (2)
•
DIL switch for sensitivity adjustment (3)
•
Simulation key (4)
Note:
For test purposes, immerse the tuning fork of VEGASWING 63 always
in liquids. Do not test the function of VEGASWING 63 with your hand.
This can damage the sensor.
6.2 Adjustment elements
1
5
2
3
4
Fig. 14: Oscillator SWE60N - NAMUR output
1 Signal lamp (LED)
2 DIL switch for characteristics reversal
3 DIL switch for sensitivity adjustment
4 Simulation key
5 EMC filter element
Single-coloured control lamp for indication of the switching condition.
•
red = High current ≥ 2.6 mA
•
dark = Low current ≤ 0.6 mA
•
red (flashing) = Failure ≤ 0.6 mA
Function/Configuration
Signal lamp (1)