beautypg.com

VEGA VEGACAP 64 - two-wire User Manual

Page 25

background image

25

9 Supplement

VEGACAP 64 • - two-wire

30015-EN-130916

Suitable signal conditioning instruments VEGATOR 521, 527, 620, 621, 622
Output signal

> 4 … < 20 mA (not standardised)

Fault message

< 2.3 mA

Switching delay

Ʋ When immersed

0.7 s

Ʋ When laid bare

0.7 s

Ʋ In case of failure

1 s

Ambient conditions
Ambient temperature on the housing

-40 … +80 °C (-40 … +176 °F)

Storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions
Process pressure

Ʋ Threaded versions

-1 … 64 bar/-100 … 6400 kPa (-14.5 … 928 psig), de-

pending on the process fitting

Ʋ Flange version

-1 … 64 bar/-100 … 6400 kPa (-14.5 … 928 psig), de-

pending on the process fitting

Ʋ Flange version ≥ 3"/DN 80, plated

-0.4 … 64 bar/-40 … 6400 kPa (-5.8 … 928 psig), de-

pending on the process fitting

Process temperature VEGACAP 64 of

316L

-50 … +150 °C (-58 … +302 °F)

Process temperature (thread or flange

temperature) with temperature adapter

(option)

-50 … +200 °C (-58 … +392 °F)

1

2

3

-50°C

(-58°F)

50°C

(122°F)

40°C

(104°F)

-40°C

(-40°F)

80°C

(176°F)

0°C

(32°F)

100°C

(212°F)

150°C

(302°F)

200°C

(392°F)

Fig. 16: Ambient temperature - Process temperature
1 Process temperature

2 Ambient temperature

3 Temperature range with temperature adapter

Dielectric constant

≥ 1.5

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug

1)

1)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.