7 Maintenance and fault rectification

VEGACAP 64 • - Relay (DPDT)

30012-EN-130916

Error

Cause

Rectification

Signal lamp flash-

es red

Electronics mod-

ule has detected a

failure

Exchange the instrument or send it in

for repair

Remove the electronics module. Check the resistance between the

two plug connections.
There must no longer be a connection (high impedance). If there is

still a connection - exchange the instrument or return it for repair

1 2

3

Fig. 27: Check the resistance in the probe
1 Shielding

2 Probe

3 Ground potential

Depending on the reason for the fault and the measures taken, the

steps described in chapter "Set up" may have to be carried out again.

7.3 Exchange of the electronics module

In general, all electronics modules of series CP60 can be inter-

changed. If you want to use an electronics module with a different

signal output, you can download the corresponding operating instruc-

tions manual from our homepage under Downloads.
Proceed as follows:
1. Switch off power supply
2. Unscrew the housing cover
3. Lift the opening levers of the terminals with a screwdriver
4. Pull the connection cables out of the terminals
5. Loosen the two screws with a screw driver (Torx size T10 or slot

4)