3 Product description

VEGACAL 62 • 4 … 20 mA/HART - two-wire

30024-EN-130917

2

1

11

13

12

14

15

10

9

3

6

4
5

7

8

Fig. 2: Layout of the type label (example)
1 Instrument type

2 Product code

3 Approvals

4 Process and ambient temperature, process pressure

5 Power supply and signal output, electronics

6 Protection rating

7 Probe length

8 Order number

9 Serial number of the instrument

10 Material, wetted parts

11 Symbol of the device protection class

12 Reminder to observe the instrument documentation

13 ID numbers, instrument documentation

14 Notified authority for CE marking

15 Approval directives

With the serial number, you can access the delivery data of the instru-

ment via www.vega.com, "VEGA Tools" and "serial number search".

In addition to the type label outside, you can also find the serial num-

ber on the inside of the instrument.

3.2 Principle of operation

VEGACAL 62 is a level sensor with partly insulated probe for continu-

ous level measurement.
The electronics functions according to the admittance principle

(phase-selective admittance processing).
It is designed for industrial use in all areas of process technology and

can be applied in all areas of industrial process measurement.
Partly insulated probes such as VEGACAL 62 are preferably used in

bulk solids.
However, the probe can also be used in non-conductive liquids such

as e.g. oil.

Probe, measured product and vessel wall form an electrical capacitor.

The capacitance is influenced by three main factors.

Area of application

Functional principle