VEGACAL 66 • Profibus PA

30037-EN-130917

Sensor address

126 (default setting)

Current value

10 mA, ±0.5 mA

Damping (63 % of the input variable)

0 … 999 s, adjustable

Input variable
Measured variable

Level of conductive liquids and solids

Measuring principle

phase-selective admittance processing (PSA)

Measuring range

0 … 3000 pF

Measuring frequency

270 kHz

Accuracy (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1

Ʋ Temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Temperature error

Ʋ < 120 pF

< 1 pF

Ʋ > 120 pF

1 % of the current measured value

Linearity error

< 0.25 % of the complete measuring range

Ambient conditions
Ambient, storage and transport tempera-

ture

-40 … +80 °C (-40 … +176 °F)

Process conditions
Process pressure

-1 … 40 bar/-100 … 4000 kPa (-14.5 … 580 psig)

Process temperature VEGACAL 66 of

316L

-50 … +150 °C (-58 … +302 °F)

1

2

-50˚C

(-58˚F)

50˚C

(122˚F)

40˚C

(104˚F)

-40˚C

(-40˚F)

80˚C

(176˚F)

0˚C

(32˚F)

100˚C

(212˚F)

150˚C

(302˚F)

Fig. 24: Ambient temperature - Process temperature
1 Process temperature

2 Ambient temperature

Dielectric constant

≥ 1.5