6 set up, 1 indication of the switching status, 2 simulation – VEGA VEGASWING 51 - contactless electronic switch User Manual
Page 19: 6set up
6
Set up
6
.1 Indication of the switching status
The switching status of the electronics can be checked via the signal
lamp integrated in the upper part of the housing.
6
.2 Simulation
The VEGASWING 51 has an integrated function for simulation of the
output signal which can be activated magnetically. Please proceed as
follows:
à Hold the test magnet (accessory) against the circle symbol with
the label "TEST" on the instrument housing
Fig. 11: Simulation of the output signal
The test magnet changes the current switching condition of the
instrument. You can check the change on the signal lamp. Please note
that the connected instruments are activated during the simulation.
If VEGASWING 51 does not switch over after several tests with the
test magnet, you have to check the plug connection and the
connection cable and try it again. If there is no switching function, the
electronics will be defective. In this case you have to exchange the
electronics or return the instrument to our repair department.
Caution:
It is absolutely necessary that you remove the test magnet after the
simulation from the instrument housing.
VEGASWING
51 • - contactless electronic switch
19
6 Set up
30212
-
EN
-111006