VEGAPULS 43 – Profibus PA

49

Technical data

Characteristics

1)

(typical values under reference conditions, all statements relate to the nominal measuring
range)

Min. span between

full and empty

> 10 mm (recommended > 50 mm)

Frequency

26 GHz

Intervals

- two-wire sensor

0.6 s

- four-wire sensor

0.5 s

Beam angle (at -3 dB)

- DN 50, ANSI 2“

15°

- DN 80, ANSI 3“

10°

- DN 100, ANSI 4“

10°

- DN 150, ANSI 6“

10°

- Clamp 2“

15°

- Clamp 3“

10°

- Clamp 4“

10°

Adjustment time (response time)

2)

> 1 s (depending on parameter setting)

Influence of the process temperature

at 0 bar not measurable;
at 5 bar 0.004 %/10 °K;
at 40 bar 0.03 %/10 °K

Influence of the process pressure

0.0265 %/bar

Emitted radar power (average)

0.717 µW

Received average emitted power

3)

- distance 1 m

0.4 … 3.2 nW per cm² (0.4 … 3.2 x 10

-9

W/cm²)

- distance 5 m

0.02 … 0.13 nW per cm²

Ambient conditions

Vessel pressure

-100 … 1600 kPa (-1 … 16 bar)

Ambient temperature on the housing

- 4 … 20 mA two-wire sensor

-40°C … +80°C

- 4 … 20 mA four-wire sensor

-40°C … +80°C

- 4 … 20 mA four-wire sensor Ex d ia

-40°C … +60°C

Process temperature (flange temp.)

-40°C … +150°C

Storage and transport temperature

-60°C … +80°C

Protection

IP 66 and IP 67

Protection class

- two-wire sensor

II

- four-wire sensor

I

Overvoltage category

III

1)

Similar to DIN 16 086, reference conditions acc. to IEC 770, e.g.
temperature 15 °C … 35 °C; moisture 45 % … 75 %; pressure 860 mbar … 1060 mbar

2)

The adjustment time (also actuating time, response time or adjustment period) is the time required by the
sensor to output the correct level (with max. 10% deviation) after a sudden level change.

3)

Average emitted power (electromagnetic energy) received by a body per cm² directly in front of the antenna.
The received emitted power depends on the antenna version and the distance.