5 Set-up
Example: requested indication 10.0 … 90.0
Input signals 0 … 20 mA; 0 … 5 V; 0 … 10 V
• remove front pane and front frame (see section
"Procedure to insert the label")
• switch on the mains
• feed the measuring signal acc. to version 20 mA or
5 V / 10 V
• set slide switch for zero point correction in position "0"
• adjust "80.0" by means of the potentiometer P
max.
(measuring range final value)
• acc. to the input signal feed initial value 0 V or 0 mA
• set slide switch for zero point correction in position
"Offset"
• adjust "10.0" by means of potentiometer P
0
.
Input signals 4 … 20 mA; 1 … 5 V; 2 … 10 V
• remove front pane and front frame (see section
"Procedure to insert the label")
• switch on the mains
• feed the measuring signal (difference) acc. to
version16 mA, 4 V or 8 V
• set slide switch for zero point connection in position "0"
• adjust "80.0" by means of the potentiometer P
max.
(measuring range final value)
• acc. to the input signal feed initial value 4 mA, 1 V or
2 V
• set slide switch for zero point correction in position
"Offset"
• adjust "10.0" by means of potentiometer P
0
.
Adjustment of the decimal point
The decimal point can be adjusted by means of a bridge
(in the example 1000).
Bridge for decimal point
Bridge plugged:
indication 100.0
indication 10.00
indication 1.000
Bridge unplugged:
indication 1000
1 0 0. 0