beautypg.com

4supplement, 3 operation, 1 fault signals – VEGA VEGATROL 20 Ex User Manual

Page 17

background image

VEGATROL 20 Ex

17

4 Supplement

3.3

Operation

If in case of overfilling, the tuning fork is covered up to the
switch point with the liquid gas, the level relay de-ener-
gizes with a delay of approx. 500 msec. and the respective
LED lights.

This remains as long as the fork is covered.

If the level reduces below the switch point, the relay
energizes with a delay of approx. 1 sec. and the
respective LED extinguishes.

4

Supplement

4.1

Fault signals

There are two error classes concerning the condition of
the LED in the front panel:
-

error during set-up

-

error detected during the self-monitoring.

Level and fail safe relays de-energize in both cases.

In case of failures during set-up, start again (switch off the
voltage and switch on again) and repeat the set-up acc. to
section 3.2.

If an error is detected, the voltage must be switched off
first, the failure (e.g. line break) removed and then the
voltage switched on again. It is not necessary to repeat
the set-up.

Error during set-up

lights

lights for approx. 2 sec.

lights

Failure reason

Failure treatment

External failure

Remove the fault and

(line break, short-circuit,

start again

fault SWING-electronic)

Set-up was carried out

Start again and repeat

when tuning fork was

set-up with uncovered

covered

tuning fork

Set-up key was pushed

Start again and repeat

too long (> 6 secs.)

set-up correctly

Error during self-monitoring

lights

lights

Failure reason

Failure treatment

Line break,

Remove failure and start

short-circuit

again

Component failure

Inform VEGA-service or
return the instrument for
invertigation

VEGATOR 526 W Ex

on

VEGATOR 526 W Ex

on