beautypg.com

5 instrument modification, 1 exchange of adjustment modules – VEGA VEGADIS 10 User Manual

Page 7

background image

VEGADIS 10

7

Instrument modification

5 Instrument modification

5.1 Exchange of adjustment
modules

Removal of an adjustment module
1

Separate VEGABAR from voltage supply
(if VEGADIS 10 i s already connected to
VEGABAR)

2

Loosen screws on cover of VEGADIS 10

3

Loosen connection cable from the te rmi-
nals

4

Loosen screws of the adjustment module

5

Detach the adjustment module and re-
move the plug connection

Addition of an adjustment module
6

Plug the plug connection of the new ad-
justment module to the plug-in socket

7

Screw on the new adjustment module

8

Connect the connection cables to termi-
nals

9

Close cover of VEGADIS 10

10 If necessar y, reconnect VEGABAR to

voltage supply

Indicating modules

Retrofitting of an indicating module
1

Separate VEGABAR from voltage supply
(if VEGADIS 10 i s already connected to
VEGABAR)

2

Loosen the screws on the cover of
VEGADIS 10

3

Remove cover

4

Connect indicating module to the screw
terminals via the four cables (see "3.2
Connection plan“)

Note:
Step 4 will be easier if the indicating module
is temporarily loosely fastened (displaced)
using two screws.

5

Place the indicating module on VEGADIS
10 in the correct position

6

Tighten the screws of the cover

7

If nec essary, reconnect VEGA BAR to
voltage supply

Indicating
module with
LC display

VEGADIS 10

Module without
adjustment

Module for
menu-guided
adjustment

Module for
basic
functions

Adjustment modules

The modular construction of VEGADIS 10
ensures that adjustment modules a nd the
indicating module can be retrofitted, ex-
changed or removed.

The connection of the modules is made:
- on the adjustment module via 4-pole plug

connection

- on the indicating module via screw termi-

nals.