Power supply

20 … 250 V AC; 48 … 62 Hz
20 … 90 V DC

Power consumption at U

N

approx. 3 VA

Measuring signal

Connection line

2-wire, unscreened (standard cable)

input

Resistance per conductor

max. 50 Ohms

Linearity error

0,1 % measuring range

Temperature error

0,01 % measuring range / 10 K

Adjustment data

Empty adjustment range

3,5 … 17 mA

Current difference between
empty and full adjustment:

> 600

µ

A

Indication

analog indication 0 … 100 %

Current output

Range

0 … 20 mA or 4 … 20 mA
(adjustable)

Load

max. 750 Ohm

Resolution

0,05 %

Linearity error

0,025 % measuring range

Temperature error

0,025 % measuring range / 10 K

Voltage output

Range

0 … 10 V, max. 1 mA (short-circuit protected)

Resolution

0,05 %

Linearity error

0,025 % measuring range

Temperature error

0,025 % measuring range / 10 K

Relay output
(only VEGAMET 308)

Contact

1 spdt

Function

two-point switching

Mode

A/B

Relay data:
Contact material

AgCdO and Au plated

min.

turn-on voltage

10 mV

switching current

10

µ

A

max.

turn-on voltage

AC 250 V, DC 60 V

switching current

AC 2A, DC 1 A

max.

breaking power

125 VA, 60 W

Permissible ambient temperature (housing)

–25

°

C … +60

°

C / –4

°

F … +140

°

F

Storage and transport temperature

–25

°

C … +70

°

C / -4

°

F … +158

°

F

Housing

Type

plastic ABS

Protection

IP 40

Dimensions

w = 96 mm, d = 76 mm, h = 102 mm

Pluggable multipoint connector (cross section area of conductor)

max. 2,5 mm

2

Weight

approx. 270 g

Technical data

3

VEGAMET 307 / 308