7 Tighten the terminal screws (2) alternately. Make sure that

the gap between clamps and pressure screw is in parallel.

Tighten the terminal screws up to a torque of 4 ±1 Nm

(3 ±0.7 lbf ft).
Hence the clamps (11) are pressed against the tube and fix

the tube of the sensor in this position

8 Fasten the locking angle (3) with the hold screw (5) and the

spring ring (4) laterally on the pressure screw (10)
Hence the pressure screw and the terminal screws are

protected against unintentional loosening

Lock fitting ARV-VB63.B - – pressure range -1 … 25 bar

7

Mounting

20036

-EN

-050804