Click in the window "Conditioning“ to

Linearisation

.

In the window "Span“ you first choose in
which unit the adjustment should be carried
out (mbar, psi, kPa). Enter if you want to edit
the percentage or the current value. If you
want to edit the percentage value, enter the
value (e.g. 80 %). Then click to

Save

. The

span adjustment is finished.

Sensor parameter adjustment – Lin-
earisation curves

Click in the window "Instrument data param-
eter adjustment“ to

Conditioning

.

In the window "Zero adjustment“ you choose
in which unit the adjustment should be car-
ried out (mbar, psi, kPa). Then click to

Save

.

The Zero adjustment is finished.

Setup

Click to

Zero

or

Span

to carry out the

respective adjustment.

Note:

The vessel or the cable must be unpressu-
rised for the zero adjustment! For the span
adjustment, it should be possible to generate
a defined pressure (e.g. 80 %) in the vessel
or on the cable.