3 technical data – VEGA VEGABAR 20 User Manual
Page 5

VEGABAR 20
5
Product description
1.3 Technical data
Mechanical data
Materials, wetted parts
Process connection
brass 2.040, stainless steel 1.4571
Diaphragm
ceramic (99,5 % Oxydeceramic)
Measuring cell seal
Viton, Kalrez
Materials, non-wetted parts
Housing
high resistance plastic PBT (Polyester)
Earth terminal
stainless steel 1.4305
Indicating module window
glass
Weight
VEGABAR
approx. 0,8 kg
Adjustment and indicating elements
Adjustment of the basic functions
2 keys, 1 rotating switch
Menu-guided adjustment with
additional function
- adjustment elements
4 keys
- indicating elements
DOT-matrix display, 3 lines with 7 figures each
Indicating module
LC-display with
- bargraph (20 segments)
- digital value (4-digits)
- tendency indicators for raising or lowering
values
Electrical data
Connection
Cable entry
Pg 13,5 (for cable-ø 5 … 10 mm)
Screw terminals
for conductor cross-sections of up to
2,5 mm
2
Protective measures
Protection
1)
IP 65
Protection class
III
Overvoltage category
III
CE-conformity
VEGABAR 20 corresponds to the requirements of EMC (89/336/EWG) and NSR
(73/23/EWG). The conformity has been judged acc. to the following standards:
EMC
Emission
EN 50 081 - 1: 1992
Susceptibility
EN 50 082 - 2: 1995
NSR
EN 61 010 - 1: 1993
1)
Use of a suitable seal for the cable in the PG is necessary for maintaining the housing protection. If the seal
used does not fit the cable, it should be replaced by a suitable one.