Density Determination Kit

5.2

Performing the density determination of liquids

Note: These instructions explain how to work with the density determination kit. They describe the procedure for
performing a density determination manually.

If you need information about operating your balance, please refer to the instruction manual which you received
with the balance.

For the density determination of liquids, use the glass beaker (1) and the optional sinker (8).

Prepare the balance for the density determination (install the platform and
bracket) as described in Section 2.

Place the empty beaker (1) on the platform and suspend the thermometer
supplied (7) from the edge of the beaker.

Suspend the sinker (8) from the bracket and ensure that it does not touch
either the beaker or the thermometer.

Tare the balance.