Calibration

For best results, calibrate the scale at regular intervals. This is especially
important if the scale is in use for prolonged periods.

• Pressing and holding the Cal button starts the

calibration process. The display should show CAL.

• The calibration process can be aborted by turning the

scale off.

• Press the On/Off Zero button to capture 0. The

display shows -C- while the scale stores the the zero
load signal.

• The display will show C 100 where 100 is the

calibration weight in grams.

• Place the appropriate calibration weight on the

platform.

• Press On/Off Zero button.

• The display shows -C- while the scale stores the

calibration point signal.

• After span capture, the display returns to the

normal weighing mode.

• The message AL E then E will appear if the

calibration steps are not followed or the wrong weight
was used.

Two Button Keypad

On/Off Zero: Pressing this button turns on the scale. This same button

operates the zero feature when the scale is on. Pressing and holding this
button turns the scale off.

Cal: Press and hold this button to begin the calibration process.