Chapter 2: Control Testing

Why Run Control Tests

Running a control test lets you know that the meter and test strips are working properly. You should run a
control test when:

You open a new box of test strips

You left the test strip container open or you think
your strips have been damaged

Your test strips were stored in extreme
temperatures or humidity

You want to check if the meter and test strips are
working properly

You drop the meter

Your test result doesn’t agree with how you feel

You want to check if you are testing correctly

About the Control Solution

Make sure you use the control solution that
matches your test strips (for example, ACCU-CHEK
Comfort Curve control solution (blue liquid) with
ACCU-CHEK Comfort Curve test strips).

When you first open the bottle, write the date on
the label – the control solution is good for three
months from that date or until the expiration date
on the bottle, whichever comes first.

Never use control solution that has expired.

Control solution can stain clothes. If you spill it,
wash your clothes with soap and water.

Always close the control solution bottle tightly.

Store at room temperature less than 90° F
(32° C). Do not freeze.