beautypg.com

Abpm 6100 hardware manual – Welch Allyn ABPM 6100 User Manual

Page 29

background image

ABPM 6100 Hardware Manual

DIR 80012720 Ver. B

29 / 30


Electromagnetic Immunity

The ABPM6100 is intended for use in the electromagnetic environment specified below. The customer or the user

of the ABPM6100 should assure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance level

Electromagnetic environment

guidance

Portable and mobile RF communications
equipment should be used no closer to any
part of the ABPM6100, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.

Recommended separation distance

Conducted RF

IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz


n.a.

P

d

= 2

.

1

Radiated RF

IEC 61000-4-3

3 V/m

80 MHz to 2,5 GHz

3 V/m

P

d

= 2

.

1

80 to 800 MHz

P

d

= 3

.

2

800 MHz to 2,5 GHz

where P is the maximum output power
rating of the transmitter in watts (W) and d
is the recommended separation distance in
metres (m).

Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey

a

,

should be less than the

compliance level in each frequency range.

b

Interference may occur in the vicinity of
equipment marked with the following

symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption

and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the ABPM6100 is used
exceeds the applicable RF compliance level above, the ABPM6100 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the ABPM6100.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.