beautypg.com

Technical characteristics, Device performance, Physical characteristics – Tyco GoodKnight 420G User Manual

Page 27: Transport and storage conditions, Conditions of use

background image

Revision B

GoodKnight 420G - Patient Manual

21

TECHNICAL CHARACTERISTICS

Device performance

Pressure range of the GoodKnight 420G: 4

cmH

2

O to 20 cmH

2

O

± 1 cmH

2

O

adjustable by increments of 0.5 cmH

2

O

Maximum achievable pressure at the patient connection
port under single fault condition:

24 cmH

2

O

Ramp time:

0 to 30 minutes

± 1 minute

device adjustable by 5-minute increments

Sound level measured in accordance with:

25.1 dBA

PrEN ISO 17510-1 :2001
Maximum sound level:

31 dBA at 10 cmH

2

O

Lifetime of the battery soldered onto the board:

> 5 years

Physical characteristics

Dimensions (W x H x L ):

5.6 x 2.9 x 7.7 in (without power supply)

Weight:

1.54 lb (without power supply module)

Transport and storage conditions

Temperature:

-20°C to +60°C (-4°F to +140°F)

Relative humidity:

up to 95 % without condensation

Relative pressure range:

500 hPa to 1060 hPa

(1 hecto-Pascal

1 cmH

2

O)

Conditions of use

Temperature:

+5°C to +40°C (+41°F to +104°F)

Relative humidity:

between 10 % and 95 % without condensation

Relative pressure range:

700 hPa to 1060 hPa

(1 hecto-Pascal

1 cmH

2

O)

Operating Altitude:

0 to approximately 7,800 ft (0-2400 m)

Disposal of the GoodKnight 420G at the end of its life

In order to protect the environment, various parts of the equipment and its accessories must be
disposed of in conformity with the local regulations.