Teledyne TED 191 User Manual

Page 21

background image

21

TELEDYNE ELECTRONIC TECHNOLOGIES

Analytical Instruments

T

ECHNICAL

S

PECIFICATIONS

SENSOR RESPONSE
TIME

Less than 6 seconds to 90% of final measured value in response to
a step change in the actual oxygen concentration being measured.

SERVICE

Technical manual will be supplied on request to user’s appropriately
qualified personnel for repair of components designated as
repairable by the manufacturer.

No parts (other than the sensor cell, and dry cell batteries)
deteriorate during normal use.

STORAGE AND
TRANSPORT

Can be stored and transported between -30°C and +50°C

Can be stored and transported at up to 75% relative humidity (non-
condensing)

GENERAL

Operation of instrument is unaffected by :

Humidity (water vapour saturated gas);

Pressure (up to 100cm H

2

O);

Nitrous oxide (N

2

O), Halothane, Enflurane or any other

anaesthetic gas or vaporised agent.

Multiple mounting bracket configurations to suit any mounting
requirement. (Custom engineered solutions available on request).