Classification – Tyco 1928L User Manual
Page 7
Classification
With respect to electrical shock, fire in accordance with UL60601-1 and CAN/CSA C22.2 No.
60601-1
This monitor is a Class I (GROUNDED) DEVICE.
These displays are classified NO APPLIED PARTS EQUIPMENT.
Protection against harmful ingress of water:
INGRESS PROTECTION (IPX1)
This monitor shall be classified as ORDINARY EQUIPMENT, not intended or evaluated for use
in the presence of flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Mode of Operation: CONTINUOUS OPERATION.
Environmental conditions for transport and storage
Temp.
Operating
0
o
C to 40
o
C
Storage / Transportation
-20
o
C to +60
o
C
Humidity (non-condensing)
Operating
30% to 70%
Storage / Transportation
10% to 90%
Altitude
Operating
1060hpa.
Storage / Transportation
0 to 40,000ft(12,192m)
Equivalent to 1013-303 hP.A
(14.7 to 4.4 psia)
For full Product Specifications refer to Appendix C
vii