TIC-LF232A

Field Intelligent Device Series

Sanitary Electromagnetic Flowmeter

Introduction

The electromagnetic flowmeter uses Faraday’s Law of

electromagnetic induction to measure the process flow.

The device consists of two units: a detector, through

which

the

fluid

to

be

measured

flows

and

in

which

low-level signals proportional to flow rates are

obtained; and a converter, which supplies excitation

current to the

detector,

and

amplifies

the

signals

from

the

detector and

then

processes

and

converts

the

signals

into

the

4–20 mA

dc current

signal.

The LF232 offers various outputs such as totalization

output and multi-range selection outputs. With it

patented Noise-Sentry original noise-suppression

circuit and advanced algorithms. The LF232 is highly

resistant to noise and provides a stable output even

when measuring fluids contain slurries. The

LF232 can

be

used

with

a

separately

mounted

detector

such

as the

LF150 or LF470. The LF232 can also be used with

conventional models of detectors.

The AF900 hand-held terminal (HART*

1

communicator) can be used to communicate with the

flowmeter from a remote place. See the

Communications signal specification for details about

HART protocol.

*1: HART protocol (Highway Addressable Remote

Transducer) is a communication protocol for industrial

sensors recommended by the HCF

(HART

Communication

Foundation).

Specifications

Overall Specifications

Measurement range in terms of flow velocity:

Conbination LF150, LF470

0 –1.0 ft/s to 0 – 32.8 ft/s (0 – 0.3 m/s to 0 –10 m/s)

Fluid conductivity: 5µS/cm or more

Ambient temperature: -4 to 140 °F (-20 to 60 °C)

(Storage temperature: -13 to 149 °F (-25 to 65 °C))

Structure: IP67 and NEMA 4X Watertight









Figure 1. LF232 Converter

Accuracy: the accuracy is dependent on the type of

detector combined with the LF232. See the following
tables for accuracy when combined with specified
detectors.

Detector combined

LF470 (Meter size 1/10", 1/6”, 1/4"

(

2.5 to 6 mm))

Accuracy

Flow rate as a

percent of range

0.3–1.0m/s 1.0–10m/s

0~50%

±

0.8%FS

±

0.4%FS

50~100%

±

0.8%FS

±

0.8% of rate

Detector combined

LF150 (Meter size 28” to 120” (700 to 3000 mm))

Accuracy: ±0.5 % of Rate*

* This pulse output error result is established under

standard operating conditions at Toshiba’s flow
calibration facility, Fuchu Japan.

* Individual meter’s measurement error may vary up

to ± 0.8 % of Rate at 3.28 ft/s (1.0 m/s) or more and
± 0.4% of Rate ±0.157 inch/s (4mm/s) at 3.28 ft/s
(1.0 m/s) or less.

* Current output: plus ± 8µA (0.05% of span).
* Refer to individual calibration data for each

individual meter’s measurement error.

LF232

Field Intelligent Device Series

Electromagnetic Flowmeter Converter