Toshiba Electromagnetic Flowmeter LF470/LF612 User Manual
Introduction, Specifications, Electromagnetic flowmeter
TIC-LF470B
Field Intelligent Device Series
Electromagnetic Flowmeter
LF470/LF612
1/10",1/6”,1/4" (2.5,4,6mm )
Introduction
The electromagnetic flowmeter uses Faraday’s Law of
electromagnetic induction to measure the process flow.
The device consists of two units: a detector, through
which
the
fluid
to
be
measured
flows
and
in
which
low-level signals proportional to flow rates are obtained;
and a converter, which supplies excitation current to the
detector,
and
amplifies
the
signals
from
the
detector and
then
processes
and
converts
the
signals
into
the
4–20
mA
dc current
signal. The LF470 is a small meter size
detector designed to measure a small amount of fluids
containing substances such as chemicals.
The wetted materials for the LF470 are corrosion
resistant ceramic and platinum electrods and are
applicable to almost any kind of fluids. The LF470 is a
lightweight palm-sized detector.Combined with
multi-functional converter LF612 (separate type) and
LF232*
1
(separate type) equipped with its patented
Noise-Sentry original noise-suppression circuit and
advanced algorithms. The LF470 is highly resistant to
noise and provides a stable output even for fluids
containing slurries. IR (Infrared) switches enable
parameter setting of the converter without removing the
cover. Flow direction can be set in either way, and its
unique 128 x 128 dot matrix LCD display allows the
LCD to be rotated electronically to be rotated to 90, 180
and 270 degrees without opening the cover
The AF900 hand-held terminal (HART*
2
communicator)
can be used to communicate with the flowmeter from a
remote place. PROFIBUS-PA*
3
interface is available
as an option.
*1: Please refer to the document ”TIC-LF232”.
*2: HART protocol (Highway Addressable Remote
Transducer) is a communication protocol for industrial
sensors recommended by the HCF
(HART
Communication
Foundation).
*3: PROFIBUS is the communication protocol for factory
and process automation that the PROFIBUS Organization
recommends. Instead of analog control with a
conventional analog signal (4-20mA), it is the fieldbus
which digitizes all signals. Flowmeters support
PROFIBUS-PA.
Figure 2. Electromagnetic Flowmeter
LF470/LF612
Specifications
Overall Specifications
Measurement range in terms of flow velocity:
0–1.0ft/s to 0–32.8ft/s (0–0.3 m/s to 0–10 m/s )
Accuracy: The overall accuracy combined with the
LF612 or LF232 converter is shown in the
following table.
Accuracy
Flow rate as a
percent of range
1.0-3.3ft/s
(0.3-1.0m/s )
3.3-32.8ft/s
(1.0-10m/s )
50 100
0.8% of FS
0.8% of rate
0 50
0.8% of FS
0.4% of FS
Note: The accuracy is measured under standard
operating conditions at Toshiba's calibration facility.
Fluid conductivity: 50µS/cm minimum
Fluid temperature:
Pipe connection material Fluid
temperature
Stainless steel and other metals
14 to 248 °F
( 10
120 )
Polyvinyl chloride
(shock-resistant)
14 to 140 °F
( 10
60 )
Ambient temperature: 14
to
140
°F (–10 to +60
°C
)
Structure: IP67
Power consumption:
・
When combined with the LF612 converter:
approximately 12W (20VA)
・When combined with the LF232 converter:
approximately 14W (25VA)
Fluid to be
measured
Signal cable
Excitation cable
Power supply
4-20mA dc output
and pulse output
Figure 1. Configuration Diagram
LF470
LF612
Certification number
Z01207
LF612 or LF232
Converter
LF470
Detector