Page 10

For digital probe matching within the instrument (version 2.1 only) additional
commands are used as follows:

CORRECTIONS

A

(turns OFF B corrections)

Turns A on

B

(turns OFF A corrections)

Turns B on

AB

(turns ON BOTH A & B corrections)

OFF (turns OFF BOTH A & B corrections)


Whatever is set by this command is retained during power-down.

LA

Lists Ch A corrections

LB

Lists Ch B corrections

CA

Enter Ch A corrections

CB

Enter Ch B corrections

Query

Says “Not enabled”, “A”, “B”, “AB”

XX*

Disables corrections

ENABLE*

Enables corrections

IC

Initialise corrections (i.e.Erase) – useful after installing software


*For these commands to be actioned the SPANs

s key must be pressed as the

command is typed.


Corrections entered as Follows

Type

TM Sends

CA

How Many Points

2

Entry=/0 Ohms

100,000

Deg C?

0.000

0100.0000 Ohms 0000.0000 Deg C ok?

Y

Entry=/1 Ohms

138.510

Deg C?

100.000

0138.5100 Ohms 0100.0000 Deg C ok?

Y

Done


This would repeat for more points.
Pressing N at the ok? Prompt will allow re-entry of the point.
Once entered, type LA OR LB TO SEE/CONFIRM CONTENTS.

Rules of entry.

1) Ohmic values must always increment from point to point
2) All values must be entered to THREE decimal places i.e:- .001,0.001,10.001 are all?
ok.