Care and maintenance, Cleaning after each use – Omron COMPACT COMPRESSOR NEBULIZER NE-C30 User Manual
Page 17

Following the cleaning instructions after each use will prevent any remaining
medication in the bottle from drying resulting in the device not nebulizing
effectively and will help prevent infections.
WARNING
Wash the nebulizer parts after each use. Dry the parts immediately
after washing.
1. Remove the inhalation accessory (mouthpiece, mask, or nosepiece)
2. Disconnect the air tubing from the nebulizer.
3. Gently twist the inhalation top counterclockwise and lift to separate the
nebulizer into two sections.
4. Remove the baffle.
5. Discard remaining medication in the medication cup.
WARNING
Always dispose of any remaining medication in the
medication cup after each use. Use fresh medication
each time you use the device.
6. Rinse all the parts of
with hot tap water. You may also wash these parts in a dishwasher.
Remove the parts after the wash and rinse cycle is completed.
DO NOT wash the mask in a dishwasher.
CAUTION
Place the nebulizer pieces in a basket recommended for dishwasher use.
Place the basket on the top shelf to avoid any dama
7. Rinse the mask with hot tap water.
8. Hand dry or air dry in a clean environment
using a soft, clean, lint-free cloth.
9. Assemble the nebulizer and store the
CAUTION
CLEANING AFTER EACH USE
17
m
a
x
5
3
Care and Maintenance
3
5
NE-C30-IM-FINAL 3-13-07 3/13/07 4:21 PM Page 16