beautypg.com

Omega Engineering FPR 200 User Manual

Page 10

background image

1. Relieve pressure in the piping system.

2. Remove the retainer ring that secures the turbine access cover.

3. Remove the access cover with a pliers, taking care not to damage

the o-ring seal.

4. Pull out the turbine assembly and the shaft.

5. Inspect the shaft for things that may have wrapped around it.

6. Inspect the turbine bearing surface for wear and elongation.

Replace as necessary.

7. Clean any rust off of the magnets that may have accumulated.

8. Reassemble the unit by placing the turbine into the body cavity

with the two magnet pockets facing inward. Place the shaft into the

turbine hole and guide it into the retaining hole in the body cavity.

Lubricate the o-ring seal with some glycerin or other lubricant and

press it into the pocket of the body. Replace the retaining ring

securely before applying pressure to the system.

Circuit recalibration (4-20mA version):
1. Place a milliamp meter into the current loop.

2. Turn off the flow going though the sensor. Adjust the OFFSET

control for a reading of 4mA on the milliamp meter.

3. Adjust the flow rate though the meter to full flow rate. Adjust the

SPAN control for a reading of 20mA on the milliamp meter. The
two controls are not interactive, so further adjustment should not
be necessary.

Circuit recalibration (0-5V

DC

version):

1. Place a voltmeter across the black & green wires of the sensor pigtail.

2. Turn off the flow going though the sensor. Verify a reading of less

than 50 mV

DC

on the voltmeter.

3. Adjust the flow rate though the meter to full-scale flow rate. Adjust

the SPAN control for a reading of 20mA on the milliamp meter.
The two controls are not interactive, so further adjustment should
not be necessary.

(10)