Enter Backdoor mode by typing:

!11,MW,1000,1[CR]

Unit will respond with:

!11,BackDoorEnabled: Y

Disable DAC update by typing:

!11,WRITE,4,D[CR]

Unit will respond with:

!11,DisableUpdate: D

7.3.2 Gas flow 0-5 Vdc analog output calibration

1.

Install jumpers J7A, J7B and J7C on the PC board for 0-5 Vdc output (see Table VI).

2.

Connect a certified high sensitivity multi meter set for the voltage measurement to the
pins 2 (+) and 1 (-) of the 15 pins D connector.

3.

Write 4000 counts to the DAC channel 1: !11,WRITE,1,4000[CR]

4.

Read voltage with the meter and calculate:

5.

Save FlowOutScaleV in to the EEPROM:

!11,MW,25,X[CR]

Where: X – the calculated AoutScaleV value.

7.3.3 Gas flow 4-20 mA analog output calibration

1. Install jumpers J7A, J7B and J7C on the PC board for 4-20 mA output (see Table VI).
2.

Connect a certified high sensitivity multi meter set for the current measurement to
pins 2 (+) and 1 (-) of the 15 pins D connector.

3.

Write 4000 counts to the DAC channel 1:

!11,WRITE,1,4000[CR]

4.

Read current with the meter and calculate:

5.

Write zero counts to the DAC channel 1:

!11,WRITE,1,0CR]

6.

Read offset current with the meter and calculate:

7. Save AoutScale_mA in to the EEPROM:

!11,MW,27,Y[CR]

Save AoutOffset_mA in to the EEPROM:

!11,MW,28,Z[CR]

Where: Y – the calculated AoutScale_mA value.

Z – the calculated AoutOffset_mA value.

Note: When done with the analog output calibration make sure the
DAC update is enabled and the BackDoor is closed
(see command below).