3.5.2 Calibration Status

The

TARGET

instrument keeps track of the date of last calibration for each sensor. To view the last calibration date,

press and hold the Menu button to enter the Pop-Up menu. Press the MENU button two times to move the asterisk to
the Cal Status position, then the SELECT button once to select the Calibration Status screen. The display shows
the status of up to three sensors. Press the SELECT button to view the remaining sensors. If any of the sensors are
overdue for calibration, Calibration due is displayed next to it. If a sensor has failed a calibration, Cal error is
displayed next to appropriate sensor.

N

OTE

:

OK

appears next to each sensor that has been calibrated within the last 90 days.

Field tests must still be

performed to confirm the performance status of each sensor.

SELECT

Example: Display of Calibration Status

3.5.3 Combustible Gas Display

Combustible gases vary considerably with respect to their flammable nature. Some gases are flammable at lower
concentrations than others. The

TARGET

instrument has ten flammable gases response characteristics programmed in

memory. Each gas has a different coefficient (K) that is used to compensate the instrument display:

Table 2: Programmed Flammable Gas Coefficients

Methane

Mth

K=1.00

Propane

Prp

K=1.88

Butane

Btn

K=2.04

Pentane

Pnt

K=2.30

Hexane

Hxn

K=2.35

Hydrogen

H2

K=1.24

Ethane

Eth

K=1.54

Benzene

Bnz

K=2.51

Toluene

Tln

K=2.47

Xylene

Xyl

K=2.99

The combustible sensor cannot determine which gas is present. When the combustible gas is unknown, it is
recommended to change the display to higher K (coefficient) value gas. Hexane or Xylene may not be the gases
present, but using their K value substantially increase, the instrument’s sensitivity to flammable gases.

To select a different combustible gas display, press and hold the Menu button to enter the Pop-Up menu. Press the
MENU button three times to move the asterisk to the Comb disp position. Press the SELECT button once to choose
this screen. Press the MENU button to sequence through the list shown in Table 2. The instrument automatically
changes the combustible display to the last one displayed.

SELECT

Example: Display, Combustible Gas

3.5.4 Locations

The

TARGET

instrument has the ability to store user name or locations. They are entered into the instrument using the

TARGET

Trakker software package, see section 7.0 for part number information. Set in accordance with software

manual.

POP-UP MENU

Auto-Zero Comb disp
*

Cal Stat

Location

SENSOR CAL STATUS

O2

OK

May 15 1999

CO

Calibration due

H2S OK

May 15 1999

POP-UP MENU

Auto-Zero

*

Comb disp

Cal Stat

Location

SET COMB GAS DISPLAY

Select type:
MTH(K=1.00)