Sensor selection, 1 sensor connections 3. electrical connections, Cdtx-90 – Omega CDTX-90 User Manual
Page 4

page 2
Caution:
Failure to fully open terminal jaws before removing wire may permanently damage instrument.
Wiring Procedure
1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end.
2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.
4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.
Wiring Removal Procedure
1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
2. When fully open, remove wire from terminal.
2
1
Wiring Tips:
• Do not route sensor cable in conduit containing AC power
wiring. Electrical noise may interfere with sensor signal.
• Routing sensor cable in grounded metal conduit will help
prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice
double wires outside the terminal.
• Sensor cable may not be extended over 100 ft.
Sensor Selection
Select the sensor with a range closest to your process minimum and
maximum values.
• CDCE-90-001 (0.01 cell): 0.055 to 100 μS (10 kΩ to 18 MΩ)
Use the CDCE-90-001 for all resistivity requirements
•
CDCE-90-01 (0.1 cell): 1 to 1000 μS
•
CDCE-90-1 (1.0 cell):
10 to 10,000 μS
•
CDCE-90-10 (10.0 cell): 100 to 200,000 μS
•
CDCE-90-20 (20.0 cell): 200 to 400,000 μS
(18 MΩ)
(10 kΩ)
CDCE-90-001
CDCE-90-20
Conductivity Range (µS)
CDCE-90 Series Sensor Ranges
Sensor Models
Ultra-
Pure
Rinse
Pure
CDCE-90-1
CDCE-90-10
CDCE-90-01
0.055 1 10
100
200 1,000
10,000
200,000
400,000
CDTX-90
≤ 100 ft. (30 m)
Conductivity/Resistivity
62.50 uS/cm
25.0 °C
ENTER
14
13
12
11
Shield (Sensor Gnd)
Black (Iso Gnd)
White (Temp In)
Red (Signal In)
CDTX-90-2 Terminals
CDCE-90 Series Standard and Certified Cells
14
13
12
11
Shield (Sensor Gnd)
Black (Iso Gnd)
White (Temp In)
Red (Signal In)
CDTX-90-3 Terminals
10
9
8
7
Shield (Sensor Gnd)
Black (Iso Gnd)
White (Temp In)
Red (Signal In)
CDTX-90-1 Terminals
3.1 Sensor Connections
3. Electrical
Connections