beautypg.com

Omega HHT21A User Manual

Page 5

background image

1

AVOID EXPOSURE

LASER RADIATION IS EMITTED

FROM THIS APERTURE

LASER RADIATION

AVOID DIRECT EYE EXPOSURE

PEAK POWER
WAVELENGTH

3mw
650nm

CLASS IIIA LASER PRODUCT

SPECIFICATIONS

This product is designed for indoor or outdoor use under the following
conditions (per IEC1010-1):

Speed Range:

5 RPM to 100,000 RPM

Accuracy:

± 1 RPM or .01% of reading

Resolution:

1 RPM

Display:

6-digit 0.45” high liquid crystal

Display Update:

Twice per second

On-Target Indication: Target (Bullseye) on lower left of display
Operating Distance: Range:

Up to 10 feet [3.1 m] from leading edge

of reflective tape
Angle:

Up to 45° from perpendicular to leading

edge of reflective tape

Power-On Switch:

Pistol grip trigger - may be locked on with latching
push-button for longer duration measurements
Auto hold of last measurement for 90 seconds

Optical System:

Visible laser light source

WARNING:
Do not view the laser beam directly as it
could be hazardous to the eyes. Do not
point the laser beam into another person’s
eyes. Do not view the laser beam with
telescopic devices.

Time Base:

Crystal controlled

Power Source:

4 “AA” batteries; Recharger socket provided for
optional rechargeable batteries

Recharger Input:

Max input 7.8 Vdc @ 50mA

Temperature:

41 °F [5 °C] to 104 °F [40 °C]

Humidity:

Maximum relative humidity 80% for temperature
up to 88 °F [31 °C] decreasing linearly to 50%
relative humidity at 104 °F [40 °C]

4

Remote/Internal
Switch

Internal Optics
Lens

Recharger Jack

Figure 1 HHT21A-R

with Optional Remote Sensor (HHT20-ROS)

CALIBRATION AND SELF-TEST

The HHT21A SERIES is a crystal controlled digital instrument that requires
no calibration. However, the accuracy of the instrument can be checked at
any time by aiming it at an old style fluorescent light and observing 7200 ±2
RPM. In countries with 50 Hz. power line frequency, the display will read
6000 ±2 RPM.

NOTE:

The Tachometer will not correctly read on energy-efficient
fluorescent lights.

BATTERIES

The HHT21A SERIES operates from four ‘AA’ size batteries which are
located under the hinged top of the instrument. To access these batteries,
grip the tapered cover at the optical lens end of the instrument and pull up.
The cover is hinged at the display end of the instrument and pivots fully out