If the [E-##] or [Err] error messages are displayed, the unit
will not function and must be returned to the factory for
repair. (See section on Warranty for return instruction).

The following charts summarize the fault condition, and the
associated indication.

Model OS951

[LO]

Target < -50 °F
(-45 °C)

Contact
Omega

Model OS953

[LO]

Target <0 °F (-
18 °C)

Use lower
OS950

Model OS955

[LO]

Target <186 °F
(86 °C)

Use lower
OS950

Model OS956

[LO]

Target <186 °F
(86 °C)

Use OS951 or
OS953

Low Target Temperature

Condition

Error

Code

Remedy

Measurement
Fault

High Ambient
Temperature

[HI A] Ambient > 122

°F (50 °C)

Move to
alternate
location

Low Ambient
Temperature

[LO A] Ambient < 32

°F (0 °C)

Move to
alternate
location

Low Battery

[N/A]

Battery power
low

Replace
battery

Dead Battery

[----]

Dead battery

Replace
battery

Other Errors

Condition

Error

Code

Remedy

Measurement
Fault

Processing
Error

[E-##]

or [Err}

Internal fault

Repeat
measurement,
if error
remains,
contact
Omega

8

Model OS951

[HI]

Target >550°F
(287 °C)

Use higher
OS950

Model OS953

[HI]

Target >1000
°F (538 °C)

Use higher
OS950

Model OS955

[HI]

Target >1207
°F (653 °C)

Use higher
OS950

Model OS956

[HI]

Target >1600
°F (871 °C)

Contact
Omega

High Target Temperature

Condition

Error

Code

Remedy

Measurement
Fault