If the [E-##] or [Err] error messages are displayed, the unit
will not function and must be returned to the factory for
repair. (See section on Warranty for return instruction).
The following charts summarize the fault condition, and the
associated indication.
Model OS951
[LO]
Target < -50 °F
(-45 °C)
Contact
Omega
Model OS953
[LO]
Target <0 °F (-
18 °C)
Use lower
OS950
Model OS955
[LO]
Target <186 °F
(86 °C)
Use lower
OS950
Model OS956
[LO]
Target <186 °F
(86 °C)
Use OS951 or
OS953
Low Target Temperature
Condition
Error
Code
Remedy
Measurement
Fault
High Ambient
Temperature
[HI A] Ambient > 122
°F (50 °C)
Move to
alternate
location
Low Ambient
Temperature
[LO A] Ambient < 32
°F (0 °C)
Move to
alternate
location
Low Battery
[N/A]
Battery power
low
Replace
battery
Dead Battery
[----]
Dead battery
Replace
battery
Other Errors
Condition
Error
Code
Remedy
Measurement
Fault
Processing
Error
[E-##]
or [Err}
Internal fault
Repeat
measurement,
if error
remains,
contact
Omega
8
Model OS951
[HI]
Target >550°F
(287 °C)
Use higher
OS950
Model OS953
[HI]
Target >1000
°F (538 °C)
Use higher
OS950
Model OS955
[HI]
Target >1207
°F (653 °C)
Use higher
OS950
Model OS956
[HI]
Target >1600
°F (871 °C)
Contact
Omega
High Target Temperature
Condition
Error
Code
Remedy
Measurement
Fault