Outputs:

Scalable 4-20mA, 0-20mA, 0-5V,
J or K thermocouple
10mV/°C Head Ambient signal

Cable Length

1 m (3.2 ft) standard, 3m (9.8 ft.) optional

Output Impedance (T/C output)

20 ohm

Minimum Load Impedance (mV output)

100K ohms

Maximum Loop Impedance (mA output)

500 ohms with 24 VDC power supply

Current Draw:

100 mA

Power Supply

11-26 VDC

Electrical Specifications

Raytek MID

General Specifications

Measurement Specifications

Accessories / Options*

Each standard MID package includes a sensing head, one mounting nut,
1 m (3.2 ft) of cable, die-cast housing with premounted electronics, and an
operator’s manual

• Adjustable or fixed mounting bracket

• Air purge jacket

• Air cooling/purging system

• Longer cable: 3m (9.8 ft.)*

• Rs232 or RS485 communication*

*Must be specified at time of order

Nominal Optical Specifications

80

(3.2)

60

(2.4)

25
(1.0)

mm (in)

Sensing Head

Electronics Housing

D:S is the optical resolution expressed as a ratio of the distance

to the resolution spot divided by the diameter of the spot.

Optical resolution for the MID is 2:1 and 10:1.
Nominal spot size based on 90% energy.

General Dimensions

Spectral Response:

8 to 14 microns

Optical Resolution:

2:1 or 10:1

Temperature Range:

-40 to 600°C (-40 to 1112°F); -25 to 600°C for
J-thermocouple output

System Accuracy:

±1% of reading or ±1°C, whichever is greater
@23°C±5°C (73°F±9°F). Thermocouple output
accuracy ±1% of reading or ±2.5°C, whichever is
greater @23°C±5°C

System Repeatability:

±0.5% of reading or ±0.5°C (1°F), whichever is greater

Temperature Coefficient:

0.15K per K or 0.15% per K, whichever is greater

Temperature Resolution:

0.3°C or 0.5°F

System Response Time:

150ms (95%)

Emissivity:

0.100 to 1.100 digitally adjustable increments of .001

Transmission:

0.100 to 1.100 digitally adjustable increments of .001

Signal Processing:

Peak Hold, Valley Hold, Variable Averaging Filter,
adjustable up to 998 seconds

Environmental Rating

NEMA-4 (IP 65)

Ambient Temperature Range:

Sensing head

0 to 85°C (32 to 185°F)

With air cooling

-18 to 200°C (0 to 392°F)

Electronics housing

0 to 65°C (32 to 150°F)

Storage Temperature

-18 to 85°C (0 to 185°F)

Relative Humidity

10 to 95%, non-condensing

Construction:

Sensing head

Stainless steel

Electronics housing

Zinc, die-cast

Weight:

Sensing head (w/1 m cable)

50 g (1.75 oz)