Nuvo Video Display Systems – User Manual


The sterilizable handle can be sterilized three ways:

• Cold, by using cleaners recommended for hospitals and authorized by a competent health

authority

• Steam
• Flash in either gravity displacement or pre-vacuum sterilizers


The minimum flash exposure times are shown in the following table:

Handle Cleaning and Sterilization

Sterilize cycle

Temperature

Exposure Time

Gravity displacement

270-274° F @ 30 psi (132-134° C @ 207
kPa

3 minutes

Pre-vacuum

270-274° F @30 psi (132-134° C @ 207
kPa

3 minutes


NOTE:
The sterilized handle should only be fitted immediately before use.

NOTE:
Service Life of Sterilizable Handle: Frequent sterilization causes natural degradation of the
replaceable handle. If signs of material fatigue such as cracking or discoloration occur, the handle
should be replaced.

NOTE:
If preferred, sterile handle covers can be used, as described below, instead of replacing the
sterilizable handle after each procedure.

Using Optional Sterile Handle Covers

The sterile handle accepts the use of the LiteGlove

®

1

flexible light handle cover by Devon, or

equivalent products, from Medical Action or DeRoyal

To install the handle cover, follow these steps:

1. Take the cover from its packaging.
2. Unfold the cover as needed.
3. Install the cover over the sterile handle.

1

LiteGlove

®

is a registered trademark of Devin Industries, Inc.