P E A K S E A R C H

6-Sep-00 13:12:32


Input

file:

HDPE .RD

DI

file:

HDPE .DI


Create background file:

NO

Create peak data file:

NO

Creeate smoothed data file:

NO

Create second derivative file:

NO


Start angle [°2

θ]:

0.200

End angle [°2

θ]:

0.260


Minimum peak tip width [°2

θ]:

0.04

Maximum peak tip width [°2

θ]:

1.00

Peak base width [°2

θ]:

2.00

Minimum significance:

0.75



Converty fixed slit intensities

to automatic slit intensities: NO


Esc=Menu F1=Start F5=Show Names F10=Help

FIGURE 1



13.

Adjust the parameters in the peak search that apply to problem with peak search.

For example,
ƒ

If there were too many peaks present, make the ‘minimum peak tip width’ larger,
or if there were too few peaks, make it smaller.

ƒ

If less intense peaks are not being detected make the ‘minimum significance’
lower.


14.

After making these adjustments press F1 twice to start the peak search again.


15.

Return to step 11.


16. Press ESC three times to return to the main menu then select EDIT (E) followed by

DI FILE (D).