background image

14

4. The oxygen concentration should be checked every 90 days for units without optional oxygen

sensors. The oxygen concentration should be checked every 6 months for units with optional
oxygen sensors.

10.2 CLEANING THE CABINET FILTERS

NOTE: All concentrator models have one cabinet filter located at rear side of the cabinet.

(FIG. 6)

FIGURE 6

CABINET FILTER LOCATION

1. Remove the cabinet filter and

CLEAN AT LEAST ONCE A WEEK depending on environmental

conditions.

(FIG.7)

2. Clean the cabinet filter with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.

(FIG. 7)

3. Dry the filters thoroughly before reinstallation.

FIGURE 7

REMOVE THE CABINET FILTERS


10.3 CABINET

Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.