Read the target plates on the walls (A,B) to detect the reference points.
If the readings of the target plates differ less than ± 5 mm, the
instrument needs not to be adjusted.
If more than ± 5 mm, proceed to the following (Adjustment)
procedure.

6

7

8

Wall A

Wall B

X side in 1

X side in 5

Target

Detector

Appr. 20m

Appr. 20m